to reading the User’s Manual. Please read the Accutrend. Plus User’s Manual carefully before using the meter for the first time. The Accutrend Plus meter is used. The Accutrend Plus system, one meter delivering both total cholesterol and glucose test results, can with Accutrend Plus may help employers improve employee productivity, improve .. 15 Cholestech LDX System User Manual. Inverness. View and Download Cobas Accutrend Plus user manual online. Accutrend Plus Blood Glucose Meter pdf manual download.

Author: Nezahn Kigore
Country: Colombia
Language: English (Spanish)
Genre: History
Published (Last): 4 July 2006
Pages: 229
PDF File Size: 6.29 Mb
ePub File Size: 8.10 Mb
ISBN: 208-1-31989-661-8
Downloads: 97995
Price: Free* [*Free Regsitration Required]
Uploader: Kigami

Performing a measurement Manuzl blood application outside the device: The reflectance photometric measurement is performed using test strips specific for each of these blood parameters. Use only the following items for cleaning: Insert the code strip smoothly into the test strip guide up to the stop in the direction of the printed arrows.

Cobas Accutrend Plus User Manual

When the instrument turns itself off, all test results obtained so far remain in the memory. The next time the Set button is pressed, the minutes can be set again with the M button. Moving backwards is not possible. Product Specifications Further information Product specifications Operating conditions and technical data Temperature range for Cholesterol: If maunal are measuring glucose, perform the plausibility check described on the next page.

The Accutrend Plus instrument Power To save power, the instrument powers itself off after 2 minutes unless a button is pressed or a new test strip is inserted.

Make sure the code strip belonging to these test strips is at hand unless the instrument has already been coded for this test strip lot.

  JUDO PROFIMAT PDF

COBAS ACCUTREND PLUS USER MANUAL Pdf Download.

Supplied with each test strip container. Error messages Display Error meaning and description Error: Page 49 Performing a measurement Test strips Code strips Make sure that no liquid enters the instrument.

Inserting Batteries Operating the instrument Inserting batteries Alphabetical Index Alphabetical index Alphabetical index Capillary blood Corrections can only be made by repeating the settings.

Quality control recommendation Starting the measurement 11 Close the measurement chamber flap. Page 8 Contents Instrument settings Brief overview of the instrument settings If a single measured value is stored incorrectly, one dash — will appear in the display. Page 96 Memory Special cases: Blood Collection Performing a measurement Blood collection Lance the outer side of the fingertip using the lancing device to obtain a large hanging drop of blood.

Measurement results which fall outside the meter measurement range are displayed as Hi above the measurement range or Lo below the measurement range.

accutrenr Hold the test strip with thumb and index finger so that the printed measurement parameter is facing upwards. Instrument settings Setting the date format In the first step, you set the date format the entire date flashes. In addition, you can flag measurements carried out with control liquids as controls.

Strip code does not match stored code The test strip belongs to a different test strip lot than the container which was last coded for this parameter. Acfutrend this manual Example of a display screen: Page 62 Performing a measurement Now take the test strip from the test strip container.

The accitrend of blood must be applied to the test strip immediately after lancing the fingertip. The properties of these test strips are then stored in the instrument.

  EGZAMIN GIMNAZJALNY 2009 MATEMATYCZNO PRZYRODNICZY ARKUSZ PDF

Performing a measurement Brief overview of the steps to be carried out Code the instrument once Insert the test strip Apply blood per test strip lot. Quality control recommendation Performing a control test Place the instrument on a level surface or hold it in your hand.

Open the measurement chamber flap.

Performing a measurement Checking the applied blood: We recommend that you always leave the beeper enabled. Introduction Operating conditions To ensure proper function of your Accutrend Plus instrument, aaccutrend the following guidelines: Quality control recommendation Preparing to perform a control test Get the test strip container for the required measurement e.

Therefore, take the correspond- ing safety measures, such as wearing disposable gloves while working. Close the container immediately after removing strip to protect the desiccant and strips, otherwise, the test accuttend may become unusable before their expiration date.

Performing a measurement Check that the whole display is functioning correctly to prevent misinterpretations due to defective display elements. Page 95 Memory To navigate within one memory area, use the M button. Test strip is unusable The test strip has already been used or is dirty.

Setting The Date Instrument settings Setting the date With the next three settings, you first enter the year, then the month and finally the day.